Status

The D/I (contact input) status is displayed on the screen.

1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5

D/I Status

Slot # 1
DI# Name Value Processing name
1 a1 Open Stream sequence
2 Open No assignment
3 Open No assignment
4 Open No assignment
5 Open No assignment
6 Open No assignment
7 Open No assignment
8 Open No assignment

Menu

Table

Slot#

F1

F2

F3

F4

F5

F6

Figure 5.7

Example of the D/I status screen

F1 (Menu):

Displays the Status Menu screen.

F4 (Table):

Displays the D/I Setup screen.

F6 (Slot #):

Sets a slot number.

TIP

• The Value data are refreshed automatically at a fi xed interval (1 second).
• On the initial screen, the smallest slot number among those of the D/I cards inserted is displayed.
• For the slot number setting, only slot numbers of D/I or DI/O cards can be accepted.
• Up to CH3 are displayed for DI/O cards.

If no D/I cards are inserted, the Status Menu screen displays “Not load” on the bottom.

Display Item

Name

Set Lower Limit Upper Limit

Unit

Remarks

Slot #

Slot number

1

5

Absolute slot number

DI #

D/I contact number

1

8

Relative channel number

Name

D/I name

8 single-byte alphanumeric

characters

Status

Status

Open/Close

Processing

name

Processing name

No assignment, Stream

sequence, Stream, Range

change Cal/Val, Operation

mode change, Alarm process

2nd Edition : May 11, 2012-00