<2. Installation, Piping, and Wiring>

2-19

IM 11B08A01-01E

Hex.socket set screw

Hex key wrench

Cable

Cover

Coupling

Gland

Guide ring

Double-deck packing
(Replace it with an appropriate one from the accessory kit.)

Identification mark

Lock nut

O-ring

Body

Connection threads

Lock nut

Flameproof packing adapter

Hex.socket set screw

Protection system
(Flameproof enclosure)

Figure 2.13

Procedure for mounting a cable packing adapter for a protection system

The cable packing adapters comply with the Technical Standard of the Ministry of Health, Labour
and Welfare, Japan.

Table 2.6

Size of double-deck packing for the protection system

Connection

port screw

Applicable

cable O.D.

Identifi cation mark

Recommended

torque for the

gland (N·m)

G3/4

ø8.0 to ø9.0

SFFP209 ø8.0 to ш9.0

17

ш9.0 to ø10.0

SFFP2010 ø9.0 to ш10.0

25

ш10.0 to ø11.0 SFFP2011 ø10.0 to ш11.0

20

ш11.0 to ø12.0 SFFP2012 ø11.0 to ш12.0

20

ш12.0 to ø13.0 SFFP2013 ø12.0 to ш13.0

20

ш13.0 to ø14.0 SFFP2014 ø13.0 to ш14.0

25

ш14.0 to ø15.0 SFFP2015 ø14.0 to ш15.0

20

ш15.0 to ø16.0 SFFP2016 ø15.0 to ø16.0

20

*:

When a cable O.D. falls under two categories, try both and choose the more suitable one.

Connecting cables to the electronics section

Six connection ports are provided in the electronics section. Use convenient ones.
The cable packing adapters (G3/4) and plates for sealing are mounted as standard. Remove the
plate before using the port for wiring, and keep it for future use. (Do not remove the plate for the
unused connection port.)

CAUTION

The minimum packing (for ø9.0 to ø10.0) is attached to the cable packing adapters at shipment.
Change it to an appropriate packing for the cable O.D. (See Table 2.7.)

2nd Edition : May 11, 2012-00