7.2.2

Malfunctioning Retention Time

Troubleshooting diagram for malfunctioning retention time is shown as followings.

F0303.ai

Malfunctioning retention time

Chromatogram is

malfunctioning

Yes

Yes

No

Yes

No

Yes

No

Yes

No

No

Malfunctioning chromatogram

Refer to "7.2.3 Malfunctioning Chromatogram".

Carrier gas pressure

is as specified

Set carrier gas pressure to

the specified value

Oven temperature

is as specified

Carrier gas

flow through detector vent

is as specified

Oven temperature

control failure

Leakage and cloggung in
piping inside the GC8000

Disconnect vent line pipng

outdside GC8000

Retention time is

still malfunctioning

Clogging in vent line piping

outside the GC8000

Change setting

values of gate timing

Figure 7.3

Troubleshooting Diagram for Malfunctioning Retention Time