<5. EtherLCD>

5-79

IM 11B08A01-01E

5.4.13 Peak

Assignments

1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5

Peak Assigments

(1/4)

Total assign 30 unassign 969
Strm# Name Assign Use GCM num
1 A1 10 0 1
2 A2 10 5 1
3 10 5 1
4 0 0 1
5 0 0 1
6 0 0 1
7 0 0 1
8 0 0 1

Menu

Exec

Assign

F1

F2

F3

F4

F5

F6

Figure 5.87

Example of peak assignments (1/4) screen

F1 (Menu):

Displays the Table menu screen.

F4 (Exec):

Executes confi guration change of peak table.

F6 (Assign): Sets stream numbers and assign numbers for the setting.

The setting is enabled only when all GCMs are in stop mode.
The total usage of peak table is set per stream.
When a stream set in GCM is selected, the GCM number is displayed in the GCM number area.
When a stream that is not set in GCM is selected, the GCM number area is blank.

Stream numbers and relative peak numbers are reset by executing F4 (Exec). Execution is
initialized to Off.

Display Item

Name

Set Lower Limit Upper Limit

Unit

Remarks

Total assign

Total assigned peak

0

999

Unassigned

Unassigned peak

number

0

999

Stream #

Stream number

1

31

Stream name

If any stream name or set name

exists, the table content is

directly shown.

Assignment

number

Relevant stream

peak assign number

0

999

Settable

Usage

Relevant stream

peak usage

0

999

GCM Number

1

6

2nd Edition : May 11, 2012-00