xvii

IM 11B08A01-01E

RUN

ALARM

POWER

SWITCH

OVERRIDE

POWER

PRESSURE

PURGING

OVERRIDE

LED (Green) of analysis "RUN"

LED (Red) of "ALARM"

LED (Green) of "POWER"

Electronic section
(Pressurized enclosure)

Door of electronic section

Operating Display

Electronics section (Pressurized enclosure)

Cover of electronic section

Door of isothermal oven

POWER LED (Yellow)

PRESSURE LED (Green)

PURGING LED (Red)

OVERRIDE LED (Yellow)

Status Display

Cover

Light sensor

Override switch

LABEL

Protection system
(Explosionproof
/ Flameproof enclosure)

FM-Y

TIIS
FM-X
ATEX
IECEx

Figure 5

Operation

WARNING

• When the door or cover of the pressurized enclosure is opened, use a gas detector to

check that the concentration of explosive gases in the ambient atmosphere is less than the
allowable limit.

• Before opening the door and cover of the electronics section, turn off the power and wait for

at least 25 minutes after obtaining the permission of the administrator.

• Electrostatic charge may cause an explosion hazard.
Avoid any actions that cause the generation of electrostatic charge, such as rubbing with a

dry cloth.

CAUTION

The isothermal oven is extremely hot after turning off the power immediately. Keep the protection
gas supplied for more than an hour after turning off the power. Keep hands away from the oven
components.