<6. Maintenance>

6-54

IM 11B08A01-01E

IMPORTANT

• During the purge, purge gas may fl ow into the sample return line, causing pressure

fl uctuation.

Prepare the sample return line and collection tank, if necessary.
• Prepare the sample return line so that the sample does not return.

CAUTION

The isothermal oven is extremely hot after turning off the power immediately. Keep the purging
air supplied for more than an hour after turning off the power. Keep hands away from the oven
components.

(5) Remove the four pipes connected to the fl owmeter.
(6) Loose the mounting screws at the base and remove the fl owmeter.
(7) Replace

the

fl owmeter in the reverse order of steps (5) and (6).

(8) Confi rm that there are no leaks in the piping.

<Replacement Parts>

K9805WA, K9805WC for gas samples (Gasket: Viton)
K9805WE

for liquid samples (Gasket: Tefl on)

K9805WB, K9805WD for gas samples (Gasket: Tefl on)
K9805WF

for liquid samples (Gasket: Daiel Par-Flour)

F0247.ai

Nut 1

Taper tube

Gasket

Connector

Nut 2

Screw (two)

Base

Figure 6.47

Taper Tube

When replacing only the taper tube, follow the procedure below (see Figure 6.47).
(1) Stop the operation. (See “3.3.4 Stopping operation” for this procedure).
(2) Turn off the supply of the sample.
(3) If the sample is liquid, purge the sample line with the purge gas (nitrogen gas or instrument

air).

2nd Edition : May 11, 2012-00