App.e-10, G) time/slope method, Peak waveform processing – Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 483: Concentration computation processing

Advertising
background image

<Appendix E Computation Scheme>

App.E-10

IM 11B08A01-01E

Coexistence of slope/time method with regular peak detection actions

Like the slope gating, time gating and zone gating method, the slope/time method is

performed by setting at the peak unit. Therefore, the slope/time method is considered as

one of the regular detection actions, allowing it to coexist with other detection methods in

one stream.

(g) Time/slope Method

Unlike the usual slope detection method or time gating method, time/slope method performs a

peak detection by using the time gating method for the peak on detection and the slope detection

method for the peak off.

Coexistence of time/slope method with regular peak detection actions

Like the slope gating, time gating and zone gating method, the time/slope method is

performed by setting at the peak unit. Therefore, the time/slope method is considered as

one of the regular detection actions, allowing it to coexist with other detection methods in

one stream.

Peak Waveform processing

This is the same as in peak waveform processing in Manual mode. (See Section 2.3.2.)

However, area calculation by tangent skimming or vertical separation can be specifi ed for each

peak.

Concentration Computation Processing

Concentration is computed when level processing for the last peak is completed in each stream.
Corrected area normalization or the standard method (external standard of indirect method) can

be group-specifi ed in the stream common list.

(a) External Standard (Approximation by Cubic Equation)

The external standard (approximation by cubic equation) computes concentration from the area

value based on the predefi ned coeffi cient.
The specifi ed standard peak is used for the coeffi cient α, S

b

, R, K, A and B.

The computation formula is:

C = RK(αS

i

/S

b

) (A (αS

i

/S

b

)

2

+ B(αS

i

/S

b

) + 1)

Where:

C : concentration
α

: calibration factor

S

b

: standard area (or height)

S

i

: measured area (or height)

R

: measurement range

K :

coefficient

A :

coefficient

B :

coefficient

(b) External Standard (Approximation by Linear Equation)

The external standard (approximation by linear equation) calculates a coeffi cient of the linear

equation from the area (or height) of two peaks specifi ed as standard peaks specifi ed as

standard peaks and calculates the concentration from the area (or height) measured using the

coeffi cient.
The equations are as follows.
(1) During

calibration

During calibration, if the range of the calibration coeffi cient is zero, calculate the standard

area, if specifi ed, calculates the calibration coeffi cient.

2nd Edition : May 11, 2012-00