Caution, 1 types of wiring and locations – Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 130

Advertising
background image

<2. Installation, Piping, and Wiring>

2-13

IM 11B08A01-01E

CAUTION

• Lay the signal wiring and electrical wiring in separate conduit pipes or ducts.
• Use independent grounding with a grounding resistance of 100 ohms or less.

2.3.1

Types of Wiring and Locations

The following types of wiring are required for the GC8000.
The wiring required varies with the specifi cations.

(A) Electric circuit and heater power
(B) Heater

power

(C) Contact output for system alarm 1
(D) Contact output for annunciator
(E) Analog input (4 to 20 mA)
(F) Contact input (Operation start/stop, mode-selection request, etc.)
(G) Contact output
(H) Communication wiring (RS-422 and analyzer bus)
(J) Analog output (4 to 20 mA), Analog hold output
(K) Grounding
(L) External I/O cutoff output (Power cutoff signal)
(M) Ethernet (twisted-pair cable)
(N) Ethernet

(optic

fi ber cable)

B

A

K

CD EFGH

JLMN

K

ABCDEFGH
JLMN

Protection system A

Control unit

Protection system B *

* Protection system B is equipped in some specifications.

With protection system

Without protection system

Figure 2.11

Cable connection locations

2nd Edition : May 11, 2012-00