App.d-9, App.d, Auto zero – Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 454

Advertising
background image

<Appendix D Action of External Input and Output Signals>

App.D-9

IM 11B08A01-01E

Type

Display width of chromatogram signals

Relevant chromatogram type

None

The display width is fi xed at 1000.000 mV.

Individual gain

The display width is determined for each peak

using the auto gain value specifi ed for each

individual peak.

A/O chromatogram

HMI chromatogram

ASET chromatogram

Overall gain

The display width is determined at once for all

detectors using the overall gain value specifi ed in

the detector signal setting.

A/O chromatogram

HMI chromatogram

Auto Zero

This is a function to change the zero point of chromatogram signals according to the auto zero

value.

Type

Display width of chromatogram signals

Relevant chromatogram type

Automatic

acquisition

The signal level automatically acquired at the

auto zero time (s) specifi ed in the detector signal

setting is regarded as the zero value.

A/O chromatogram

HMI chromatogram

Auto zero

value

The auto zero value specifi ed for each individual

peak in the detector signal setting is used.

A/O chromatogram

HMI chromatogram

Invalid

The zero value is fi xed at 0.000 mV.

1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5

Detector Setup

(1/2)

1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5

Detector Setup

(2/2)

Oven #

1 Det # 1 ( FID )

Oven #

1 Det # 1 ( FID )

>

- Filtering const 0.500

>

- Auto gain setting Unexecuted

- Sample rate 40 ms

- Full scale value 0

- Square root calc req * * * * *

- Auto zero Off

- Flame detect level * * * * *

- Auto zero value 0.0000 mV

- Sense set 4.0 mV

- Auto zero time 0.0 s

- Sense set 10 times

Menu

Oven#

Det#

Menu

Oven#

Det#

F1

F2

F3

F4

F5

F6

F1

F2

F3

F4

F5

F6

Figure 10 Auto Gain and Auto Zero Settings (Detector Signal Setting Screen on GC-HMI EtherLCD)

1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5

Peak Setup-Specifi c

(1/5)

Stream #1 Peak #1
Using/Assigned peak num 1 / 10
>

- Process (Type) Peak

- Process (Detail)

External 3rd

- Name H2
- Execution Excuted
- Synchronize to #01 #001
- Output fl ag Provided (31001)
- Range number 0
- Auto gain value 4

Menu

Copy1

Copy2

Strm#

Peak#

F1

F2

F3

F4

F5

F6

Figure 11 Auto Gain Value (Individual Peak Setting Screen on GC-HMI EtherLCD)

2nd Edition : May 11, 2012-00