<7. Troubleshooting>

7-4

IM 11B08A01-01E

No.

MESSAGE

Alarm Contents

Threshold Item

Threshold

Threshold

Setting

61 AI A/D ERR

AI card A/D

calibration error

Calibrate 16-bit

AD of temperature

controller signal

and AI signal upon

power-on

Error to calibrate 16-

bit AD of temperature

controller signal

and AI signal upon

power-on

N/A

62 TMP CARD1 A/D

ERR

Temp. ctrl card 1 A/D

cal. error

63 TMP CARD2 A/D

ERR

Temp. ctrl card 2 A/D

cal. error

64 TMP CARD3 A/D

ERR

Temp. ctrl card 3 A/D

cal. error

65 OVEN1 TMP A/D

ERR

Oven 1 temp. A/D

cal. error

66 OVEN2 TMP A/D

ERR

Oven 2 temp. A/D

cal. error

67 OVEN3 TMP A/D

ERR

Oven 3 temp. A/D

cal. error

71 OVEN1 COM ERR

Oven 1

communication error

Communication with

OVEN 1

Communication

between OVEN and

CPU is disconnected

for a set period of

time

N/A

72 OVEN2 COM ERR

Oven 2

communication error

Communication with

OVEN 2

73 OVEN3 COM ERR

Oven 3

communication error

Communication with

OVEN 3

74 TMP1 CTL ERR

Temp. sensor 1

temp. cntrl error

Exceeds the

deviation limit

setting for Oven

1 temperature

setting value and

measurement value.

Deviation exceeds

the value set

Deviation

temperature

setting screen

75 TMP2 CTL ERR

Temp. sensor 2

temp. cntrl error

Exceeds the

deviation limit

setting for Oven

2 temperature

setting value and

measurement value.

76 TMP3 CTL ERR

Temp. sensor 3

temp. cntrl error

Exceeds the

deviation limit

setting for Oven

3 temperature

setting value and

measurement value.

77 TMP1 AD ERR

Temp. sensor 1

standard AD error

Standard ZERO/

MIDDLE/SPAN

values exceed the

range

Each value exceeds

x

N/A

78 TMP2 AD ERR

Temp. sensor 2

standard AD error

79 TMP3 AD ERR

Temp. sensor 3

standard AD error

81 TMP1 OVER LIMIT Temp. sensor 1 over

limit

Temperature sensor

value

Upper limit: User-set

value

Lower limit: –20°C

N/A

82 TMP2 OVER LIMIT Temp. sensor 2 over

limit

83 TMP3 OVER LIMIT Temp. sensor 3 over

limit

84 TMP1 HIGH

Overtemp protection

zone1

N/A

85 TMP2 HIGH

Overtemp protection

zone2

86 TMP3 HIGH

Overtemp protection

zone3

87 TMP1 SENSOR

BURN OUT

Temp. sensor 1

failure

Temperature sensor

value

Temperature

exceeds the upper

limit of 350°C

N/A

88 TMP2 SENSOR

BURN OUT

Temp. sensor 2

failure

89 TMP3 SENSOR

BURN OUT

Temp. sensor 3

failure

2nd Edition : May 11, 2012-00