<7. Troubleshooting>

7-13

IM 11B08A01-01E

(2) FID

Troubleshooting diagram for malfunctioning chromatogram is shown as followings.

F0305.ai

Malfunctioning chromatogram

Flow rate

of combustion gas is as

specified

Yes

Yes

No

Yes

No

Yes

No

Yes

No

Yes

No

Yes

No

Yes

No

Yes

No

No

Yes

Yes

No

Turn detector

DET FLAME OUT

is occured

Status of

detector is ON

Adjust flow rate of

combustion gas

Flow rate

of combustion gas is as

specified

Yes

No

High voltage

value for signal at DET card is

as specified

Yes

No

High voltage

value for signal at DET is

as specified

Adjust flow rate of

combustion gas

Replace detector card

Replace wiring for detector

Voltage value

for ignition at DET board is

as specified

Replace detector board

Yes

No

Voltage

value for ignition at DET is

as specified

Replace wiring for detector

Replace detector

Baseline of

detector is malfunctioning

Column system failure or

sampling system failure

Temperature of oven is

stable

Wait for stabilization of

oven temperature

Combustion air is clean

Use clean combustion air

External pipng is clean

Ude clean piping

No

Replace detector

Instalation

resistance of detector is

good

Replace wiring for detector

Wiring for detector is

good

Replace detector board

Figure 7.5

Troubleshooting Diagram for Malfunctioning Chromatogram (FID)