<1. Overview>

1-41

IM 11B08A01-01E

F0121.ai

Figure 1.20

Stream sequence window on the GCM operation status screen on the GC-HMI analyzer

operation display

11/03/22

15:15:45

Stream Sequence (1/4)

GCM #1 Seq #1

- #01 Stream 1 (STR1)

- Excuted

- #02 Stream 2 (STR2)

- Excuted

- #03 Stream 3 (STR3)

- Excuted

- #04 Stream ** (********) - Not excuted
- #05 Stream ** (********) - Not excuted
- #06 Stream ** (********) - Not excuted
- #07 Stream ** (********) - Not excuted
- #08 Stream ** (********) - Not excuted
- #09 Stream ** (********) - Not excuted

Menu

Gate

Valve

Atm

GCM#

Seq#

F1

F2

F3

F4

F5

F6

Figure 1.21

An example of the stream sequence setting screen on the GC-HMI EtherLCD

● In the case where a stream sequence change command is received during the stop mode in

the stream sequence status [1A1]

The measurement status is immediately changed to the requested stream sequence status.
In this case, the operation mode remains unchanged from the stop mode.

* Requests for changes to the same stream sequence number are invalid.