11 multirange setup – Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 328

Advertising
background image

<5. EtherLCD>

5-77

IM 11B08A01-01E

Display Item

Name

Set Lower Limit Upper Limit

Unit

Remarks

GCM #
Val #

Val number

1

6

Validation

stream

0

31*

* Only val streams belonging to

GCM can be set. (Stream setup

screen )

Validation

times

0

30

Auto

validation

specifi cation

Not done, Done

5.4.11 Multirange

Setup

When using peak setup-specifi c to set a range number, the multirange setup is made effective.
When synchronizing the data in the range number with peak setup-specifi c, changes can be
made bidirectionally.

1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5

Multirange Setup

Range # 1
>

- Process (Type) Distillation

- Process (Detail)

Std B.P proc

Menu

Get

Range#

F1

F2

F3

F4

F5

F6

Figure 5.85

Example of multirange setup screen

F1 (Menu):

Displays the Table Menu screen.

F4(Get):

Gets the set values of the Process assign, measurement unit, measurement
range, etc. of the specifi ed peak.

F6 (Range #): Specifi es a range number (1 to 31).

<Range synchronization rules>

Stream type

Range synchronization

Measurement Calibration Validation

Calibration

run

Sample run

Valid

Invalid

Valid

Invalid

Invalid

Display Item

Name

Set Lower Limit Upper Limit

Unit

Remarks

Range #

Range number

1

31

Process

(Type)

See the Display Data List for

Peak setup-specifi c screen.

Process

(Detail)

See the Display Data List for

Peak setup-specifi c screen.

Measurement

Unit

See the Display Data List for

Peak setup-specifi c screen.

Measuring

range

See the Display Data List for

Peak setup-specifi c screen.

Ref peak

number

0, 100%

value

2nd Edition : May 11, 2012-00