<Appendix D Action of External Input and Output Signals>

App.D-18

IM 11B08A01-01E

1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5

D/I Setup

Slot #

1 DI # 1

>

- Signal name S1P1-R1

- Process assign Range Select
- Stream number 1 (GCM 1)
- Peak number 1
- Range number 1

Menu

Status

Slot#

DI#

F1

F2

F3

F4

F5

F6

Figure 24 Process Specifi cation: Range Switching (D/I Setting Screen on GC-HMI EtherLCD)

1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5

Peak Setup-Specifi c

(2/5)

Stream #1 Peak #1 External 3rd
- Detector number 1-1
>

- Measuring unit 1

- Measuring range 100.000
- Gate cut method Slope gate
- Integ method

Skimming

- Peak polarity Positive
- Detected slope 0.0010 mV/s

Menu

Copy1

Copy2

Strm#

Peak#

F1

F2

F3

F4

F5

F6

1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5

Multirange Setup

Range # 1
>

- Process (Type) Peak

- Process (Detail)

External 3rd

- Measuring unit 1
- Measuring range 50.000
- Ref peak number #40 #001

Menu

Get

Range#

F1

F2

F3

F4

F5

F6

Figure 25 Individual Peak Setting Screen and Multi-Range Setting Screen on GC-HMI EtherLCD

Calibration/Validation

The Calibration/Validation function of D/I is a function to give a command to change the

measurement status to the status of a specifi ed calibration or validation number using contact

input as a trigger.
Six patterns of calibration setting and validation setting can be used for each GCM.
To give a command to all GCMs, set “0” for the GCM number.