18 a/i setup – Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 338

Advertising
background image

<5. EtherLCD>

5-87

IM 11B08A01-01E

F1 (Menu):

Displays the Table menu screen.

F3 (Status):

Displays the A/O status screen.

F5 (Slot #):

Sets a Slot number.

F6 (AO #):

Sets a A/O number.

• When A/O chromatogram output is set, other items are displayed with * (asterisk).
• #1 Output stream and #2 Output stream cannot be set in the same Stream. If this setup is

attempted, the error message “Already assigned in request stream” is displayed.

• If a detector that does not exist is selected in the setting of the A/O chromatogram, the error

message “Not load the detector” is displayed.

Display Item

Name

Set Lower Limit Upper Limit Unit

Remarks

Slot#

Slot number

1

5

AO #

A/O channel number

1

8

GCMNumber

1

6

AO

Chromatogram

Normal output, Det1-1, Det1-2,

Det2-1, Det2-2, Det3-1, Det3-2

* Multiple OA channels can

be assigned to the same

detector. However, up to a

maximum of 8 channels.

#1 Output

stream

(0), 1

(99), 31

Lower Limit (0) cancels the

output of the relevant peak.

Upper Limit (99) gives output of

analyzed values of all streams.

#2 Output

stream

(0), 1

31

Lower Limit (0) cancels the

output of the relevant peak.

#1, #2 Output

peak

(0), 1

999

Lower Limit (0) cancels the

output of the relevant peak.

#1, #2 Partial

hi

0.001

1.000

#1, #2 Partial

low

0.000

0.999

5.4.18 A/I

Setup

1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5

A/I Setup

Slot #

1 AI # 1

>

- Signal name AA

- Filter const 1.000

Menu

Status

Slot#

AI#

F1

F2

F3

F4

F5

F6

Figure 5.99

Example of A/I setup screen

F1 (Menu):

Displays the Table menu screen.

F3 (Status):

Displays the A/I status screen.

F5 (Slot#):

Sets a Slot number.

F6 (AI#):

Sets an A/I number.

2nd Edition : May 11, 2012-00

5