2 operating parameters – Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 292

Advertising
background image

<5. EtherLCD>

5-41

IM 11B08A01-01E

5.4.2 Operating

Parameters

1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5

Operating Parameters

(1/8)

Oven #

1 Carrier gas # 1

>

- Carrier gas type - - -

- Carrier gas pressure

0.0 kPa

Menu

Oven#

Gas#

F1

F2

F3

F4

F5

F6

Figure 5.53

Example of the operating parameters screen (1/8)

F1 (Menu):

Displays the Table Menu screen.

F5 (Oven #): Specifi es an oven number.
F6 (Gas #):

Specifi es a carrier gas number.

TIP

The oven numbers set to Not provided on the Temperature Control Set cannot be selected.

• The initial screen displays the smallest oven number among those assigned to isothermal

ovens on the Temperature Control Set.

Display Item

Name

Set Lower Limit Upper Limit

Unit

Remarks

Oven #

Oven number

1

3

Carrier gas # Carrier gas number

1

2

Carrier gas

types

Carrier gas type

○ *

---

H

2

He

Ar

N

2

* --- indicates not used.

Carrier gas

pressure

Carrier gas pressure

0

0.0

490

72.5

kPa

psi

Numeric entry

psi (see the defi nition of SI units)

500 kPa = 72.5 psi

*:

The settings are valid after rebooting or turning off/on its power.

1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5

Operating Parameters

(2/8)

Oven #

1 Utility gas # 1

>

- Detector number - - -

- Utility gas function - - -
- Utility gas type

- - -

- Utility gas pressure 0.0 kPa
- Utility gas fl owrate

Menu

Oven#

Gas#

F1

F2

F3

F4

F5

F6

Figure 5.54

Example of the operating parameters screen (2/8)

2nd Edition : May 11, 2012-00