<6. Maintenance>

6-52

IM 11B08A01-01E

L9862AD Gasket

2

F0245.ai

Filter Element

Body

Nut

Bushing

Gasket 1

Gasket 2

Figure 6.45

(2) Atmospheric Pressure Balancing Valve and/or Needle Valve for

Controlling Flow Rate

See Figure 6.46.
(1) Stop the operation. (See “3.3.4 Stopping operation” for this procedure).
(2) Stop the supply of power.

CAUTION

The isothermal oven is extremely hot after turning off the power immediately. Keep the purging
air supplied for more than an hour after turning off the power. Keep hands away from the oven
components.

(3) Turn off the supply of the purging air.
(4) If the sample is liquid, purge the sample line with the purge gas (nitrogen gas or instrument

air).

• When the stream switching valve is a pneumatic valve, change the analyzer status to

Manual and turn on the stream switching valve on the sample line to let the purge gas fl ow
in.

• When the stream switching valve is a stop valve, close all the stop valves and open the stop

valve on the sample line to let the purge gas fl ow in.

(5) Close all the stream switching valves.

IMPORTANT

• During the purge, purge gas may fl ow into the sample return line, causing pressure

fl uctuation.

Prepare the sample return line and collection tank, if necessary.
• Prepare the sample return line so that the sample does not return.

2nd Edition : May 11, 2012-00