<5. EtherLCD>

5-101

IM 11B08A01-01E

5.4.27

MODBUS CLIENT Setup

1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5

MODBUS CLIENT setup

MODBUS # 1
>

- IP address 0.0.0.0

- Flag Disconnect

Menu

MODBUS#

F1

F2

F3

F4

F5

F6

Figure 5.114

Example of MODBUS Client Setup screen

F1 (Menu):

Displays the Table menu screen.

F6 (MODBUS):

Sets a MODBUS number.

Up to 20 GCs which enable communication between GC are able to be registered on the LCD
screen. The procedure is as follows.
1. Press F6 (Modbus #), and the setup screen is selected.
2. Set the IP address of the GC of the connection destination and change the fl ag to Connect.

Note: While the GC of the connection destination remains Connect, it cannot be re-registered on another setup screen. A message

advising that the same GC cannot be registered is displayed.

Display Item

Name

Set Lower Limit Upper Limit

Unit

Remarks

MODBUS #

MODBUS Number

1

20

IP address

○ *

Flag

○ *

Disconnect, Connect

*:

The settings are valid after rebooting or turning off/on its power.