Safety instructions – Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 21

Advertising
background image

xx

IM 11B08A01-01E

Safety Instructions

This content described as follows is only for ATEX and IECEx.

● Specifi cation of safety system

Protection system

Oven safety unit

Power supply

Install in accordance with the specifi cation

of GC8000

Ambient condition
Threshold

value

Pressure

392 ±20 Pa

Flow rate of protection gas

35 L/min

Time

Purging time

21 ±3 min

11 ±3 min

Safety response time

< 2 sec

Safety

characteristics

HFT (*1)

1

SIL (*2)

2

Proof test interval

No needed

At least once within

3 years

Replacement interval (*3) At least once within

12 years

At least once within

27 years

Type of

protection

Protection system itself

Ex d

Ex px

Protection target

Ex px

(*1)

HFT: hardware Fault Tolerance

(*2)

SIL: Safety Integrity Level

(*3)

This interval is the longest period to maintain explosion proof function and does not mean that the system is under warranty
during the interval.

● Installation and wiring works
The safety system is built-in GC8000. Installation and wiring into GC8000 are completed before
shipping.

• Do not use for other than GC8000.
• Do not change connection of cables and harnesses.

● Action in Emergency
Relays and SSRs in safety system will be in safe state (contact opens) when de-energized.
In emergency, stop supply of power to GC8000 immediately and contact your nearest Yokogawa
representative.

• In a normal situation, follow the procedure of “3.3.4 Stopping operation” for shutdown, or

GC8000 may be damaged.

● Regular proof test
In order to maintain the safety function, proof test at least once while test period below is
necessary. Contact your nearest Yokogawa representative.
If proof test is not performed for longer than test interval, GC8000 cannot be used as explosion
proof equipment.

Protection system

Oven safety unit

Test interval

No needed

3 years

● Replacement
In following case, stop supply of power to GC8000 immediately and replace the safety system.

• Safety function does not work correctly.
• Safety system is used over replacement interval.

Replacement is needed even if all safety functions work correctly.
Contact your nearest Yokogawa representative.

Protection system

Oven safety unit

Replacement interval

12 years

27 years

2nd Edition : May 11, 2012-00