Peak copy 1 – Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 322

Advertising
background image

<5. EtherLCD>

5-71

IM 11B08A01-01E

Peak copy 1

Press F2 (Copy1).

1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5

Peak Copy 1

(1/4)

Total Assign # 30
Strm# Stream name Assign Peak num
1 A1 10 1
2 A2 10 5
3 10 5
4 0 0
5 0 0
6 0 0
7 0 0
8 0 0

Menu PeakSet

Copy

F1

F2

F3

F4

F5

F6

Figure 5.79

Example of peak copy 1 (1/4) screen

F1 (Menu):

Displays the Table Menu screen.

F2 (PeakSet): Displays the Peak Setup-Specifi c screen.
F6 (Copy):

Executes copy after entering Source stream and Destination stream.

• Copy destination peak must be an independent peak.
• See Peak synchronization and peak copy items regarding data for copy.

Display Item

Name

Set Lower Limit Upper Limit

Unit

Remarks

Total Assign # Total assign

0

999

Total assigned peak of all

streams

Strm#

Stream number

1

31

Stream name
Assign

0

999

Total assigned peak of stream

Peak num

0

999

Total used peak number of

stream

<Peak copy rules 1 (Peak copy screen 1)>
1. Copy peak assigned to stream source and destination.
2. When stream source and destination of assignment number are different, use the stream

with the smaller assignment number as the standard.

3. When the assignment number of either the stream source or destination is zero, error will

occur and copy cannot be made.

2nd Edition : May 11, 2012-00