Atm action setting screen – Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 310

Advertising
background image

<5. EtherLCD>

5-59

IM 11B08A01-01E

Enter a time with a tenkey.
Pressing F1 (Seq) navigates to the SYS sequential display screen.

Only valves belonging to the relevant SYS can be set by F6 (Valve#).

Display Item

Name

Set Lower Limit Upper Limit

Unit

Remarks

GCM #

GCM number

1

6

SYS #

SYS number

1

6

Method #

Method number

1

6

Valve

Number

Valve number

1 − 1

3 − 7

0: None

1 to 3 Oven

1 to 7 Valve

1st to 3rd ON

Time

0

Analysis

cycle − 2

seconds Valve time set

Some entry data are invalid.

* When ON time is invalid, OFF

time cannot be set.

1st to 3rd

OFF Time

ON time

Cycle Time

− 2

seconds

ATM Action Setting Screen

Press F4 (AtmV).

1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5

ATM Action Setting

GCM #1 / SYS #1 / Method #1
ATM valve 1-1
>

- 1st ON time

* * * * * * * s

- 1st OFF time

* * * * * * * s

- 2nd ON time

* * * * * * * s

- 2nd OFF time

* * * * * * * s

- 3rd ON time

* * * * * * * s

- 3rd OFF time

* * * * * * * s

Seq

Gate

Valve

DO

F1

F2

F3

F4

F5

F6

Figure 5.73

Example of the ATM action setting screen

F1(Seq):

Displays SYS sequential Display screen.

F2(Gate):

Displays Gate of the action set screen.

F3(Valve):

Displays ON/OFF time screen of a valve (1st, 2nd & 3rd of maximum No. 1 to
18).

F5(DO):

Displays D/O Setup screen in the action set screen

• ATM valve number set in GCM setup is displayed for ATM valve number (display only).

Functions of the function keys F2, F3 and F5 are the same as those in SYS method setup (Part
1; Main).

2nd Edition : May 11, 2012-00

5