<Appendix C Standard Specifi cations>

App.C-5

IM 11B08A01-01E

Contact operation: Open/Close

Contact

Annunciator Output

Between NC and

COM terminals

Between NO and

COM terminals

State 1

Open

Close

State 2

Close

Open

Power supply: OFF

Open

Close

4.2.3

Analog Output (Option, Up to 4 cards)

Number of outputs:

8 per card, maximum of 32

Signal type: 4 to 20 mA DC

Isolation: Channel isolation, system isolation

(selectable)

Load: 300

Ω or less

Output Types:

Analysis results (analog hold output)/

Chromatogram output

The maximum number of outputs

is 32 analog hold outputs, and 8

chromatogram outputs even in multi-

analog output card.

Note: The analog hold output cannot coexist with GCCU

communication (but the chromatogram output can
coexist with GCCU communication).

Output range: Any setting is possible within the

measuring range. Auto gain can be set

for chromatogram output.

4.2.4

Analog Input (Option, Up to 4 cards)

Number of inputs:

4 per card, maximum of 16

Signal types: 1 to 5 V DC, 4 to 20 mA DC

Isolation: Channel isolation

Accuracy: ±0.5% of full scale (–10 to 50°C)

Input types: Analysis results (e.g. other

analyzers), temperature (e.g. ambient

temperature), pressure, etc.

Function: The following values can be output

by Modbus protocol in the DCS

communication after computing input

values.

Average value:

The one-second average of analog

values measured at every 200 msec

and fi ltered by a predetermined

constant

Current value:

The value at the time set in a cycle

time

External I/O Cutoff Output:

Number of outputs: 1

Function: Monitoring the purge air pressure

in the electronics section, applying

power (24 V DC) to the signal

interrupter when the state is normal.

Signal interrupter (Rack-mounted type:

K9806AE):

FM-X, ATEX, IECEx, TIIS:

Additionally required. Signal is

interrupted by power supply OFF

signals from the external I/O cutoff

output.

FM-Y: Not

required

Note: For installation in hazardous area, an explosionproof

structure authenticated by a relevant certifi ed body
shall be prepared by customer.

The same number unit of the signal interrupter as
input card is required.

4.2.5

Contact Output (Option, Up to 4 cards)

Number of contacts: 5 per card, maximum of 20

Function: The following settings are possible at

each contact point.

Stream sequence:

Outputs when the specifi ed stream

sequence is operated.

Stream: Outputs when the specifi ed stream is

operated.

Operation mode:

Outputs when the specifi ed operation

mode is operated.

Alarm:

System alarm:

Outputs when the system alarm

occurs.

Composition Alarm:

Outputs when the concentration

alarm or the retention time alarm

occurs.

Timing: Outputs at the time to be set.

Calibration/Validation:

Outputs when the specifi ed

calibration or validation is operated.

Stream valve selection:

Output for the external sampling

equipment, up to 31.

Stream identifying:

Output for the analog hold output, up

to 5 points per 1 GCM (5 bits, up to 31

streams)

Contact specifi cations:

SSR or Relay contact output, c-contact

(NO, NC, COM)

Contact rating:

SSR contact output:

100 to 240 V AC, max. 2 A (Load),

Relay contact output:

24 V AC, max. 2A (Load)

Contact operation:

ON/OFF operation

Contact state: Selectable from Open or Close on

operation (Open when power supply is

turned off.)

External I/O Cutoff Output:

Number of outputs: 1

Function: Monitoring the purge air pressure

in the electronics section, applying

power (24 V DC) to the signal

interrupter when the state is normal.

Signal interrupter (Rack-mounted type, AC:

K9806AN, DC:K9806AJ):

FM-X, ATEX, IECEx, TIIS:

Additionally required. Signal is

interrupted by power supply OFF

signals from the external I/O cutoff

output.

2nd Edition : May 11, 2012-00

App.C