Detector type: fid-mc – execution screen – Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 287

Advertising
background image

<5. EtherLCD>

5-36

IM 11B08A01-01E

Display Item

Name

Set Lower Limit Upper Limit

Unit

Remarks

Oven #

Oven number

1

3

Det #

Detector number

1

2

Detector type

FID-MC

Det operation

0: Off

1: On

* For this operation, the user

level C or higher is required and

the operation mode must be

Stop in Manual status.

Reset status

0: Initialized

1: Setting done

If the voltage value is zero,

Initialized is displayed;

otherwise Setting done.

Detector type: FID-MC – Execution screen

1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5

Det Adjustment

Oven #

1 Det # 1 ( FID-MC )

>

- Det operation Off

MC voltage value 0.0 V
- Target MC current

Main

<<

<

>

>>

F1

F2

F3

F4

F5

F6

Figure 5.47

Example of the detector type: FID-MC – execution screen

F2 (Main):

Displays the Det Adjustment (Reset) screen.

F3 (<<):

Decreases the value by 16 steps within the range of 0 to 255 (for the MC, by
approximately 0.75 V within the range of 10.0 V to 22.0 V).

F4 (<):

Decreases the value by 1 step within the range of 0 to 255 (for the MC, by
approximately 0.05 V within the range of 10.0 V to 22.0 V).

F5 (>):

Increases the value by 1 step within the range of 0 to 255 (for the MC, by
approximately 0.05 V within the range of 10.0 V to 22.0 V).

F6 (>>):

Increases the value by 16 steps within the range of 0 to 255 (for the MC, by
approximately 0.75 V within the range of 10.0 V to 22.0 V).

• To toggle the Det On/Off status, set the cursor on the Det operation and press the Set/Ent

key. The screen displays “Setting succeeded” or “Setting failed” in the second line on the
bottom.

• If the Up/Down command keys are pressed to adjust the value within the range of 0 to 255

and the command is successfully sent internally, “Setting succeeded” is displayed. Users
can confi rm the results with the voltage value and current value refreshed every second.

• Selecting the Target MC voltage displays Target MC voltage (AA.A/B.B):. If the value

17.5/0.5 is entered, it will be shown as 17.5+/–0.5 V, 5.0 as 5.0+/– V and 2/1 as 2.0+/– 1.0 V.

• The numbers of ovens to which no detectors are installed cannot be set via F5.

2nd Edition : May 11, 2012-00