<2. Installation, Piping, and Wiring>

2-25

IM 11B08A01-01E

Screw
for
Ground

Figure 2.21

Earth terminal for the GC8000

2.3.5

Connecting Signal Cables

CAUTION

Be sure to keep the power and signal cables apart. Avoid placing them in parallel.

NOTE

• For analog input, use twisted pair cables with a common shield (a twist pitch of 50 mm or

less), to avoid induction noise.

Use twisted pair cables for digital signals as well.
• Stranded cables are superior to single-conductor cables in the following respects:

• Stranded cables are more fl exible and easy to lay in a curved pit or cramped space.
• Stranded cables provide better contact with crimp-on terminal, with less aging over time.

• Secure the cables so they do not weight on the terminals.
• Fasten the terminal screws securely.

Signal Cable Termination

CAUTION

• Use crimp-on terminal with insulated covering.
• Wire crimp-on terminal with the dedicated tool.
• The tool must be suitable for the size of wires.

Use crimp-on terminal for all signal cables.
The specifi cations of the crimp-on terminal are determined by the nominal cross sectional area of
the power cable.