Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 57

Advertising
background image

<1. Overview>

1-30

IM 11B08A01-01E

The maximum temperature is limited in accordance with each explosion-proof standard and its
allowable temperature level and type of detector installed. The values of the setting temperature
are described in the operation manual, and can be checked on the operation condition
confi guration screen of the GC-HMI (touch panel) EtherLCD.
Refer to Figure 1.11.

The following parts are in the large isothermal oven and isothermal oven.

• Pressure regulator for carrier gas
• Restrictor
• RV
• LSV
• Column
• Detector
• Flame

arrester

• Mesh

arrester

The specifi cation determines the number and types of parts. Refer to Table 3.1 for the maximum
number of parts for each oven.

11/03/22

15:15:45

Temperature Setup

Oven #1

- Oven max temp

320.0 °C

> - Oven temp setpoint 145.0 °C
- LSV temp setpoint

150.0 °C

- FPD temp setpoint

***** °C






Menu

Menu

Menu

Menu Oven# Oven#

F1

F2

F3

F4

F5

F6

11/03/22

15:15:45

Temperature Setup

Oven #2

- Oven max temp

320.0 °C

> - Oven temp setpoint 30.0 °C
- LSV temp setpoint

***** °C

- FPD temp setpoint

***** °C






Menu

Menu

Menu

Menu Oven# Oven#

F1

F2

F3

F4

F5

F6

2nd Edition : May 11, 2012-00

Advertising