<6. Maintenance>

6-5

IM 11B08A01-01E

(a) One enclosure

(b) Two enclosures

Figure 6.2

(1) Pressure Switch for Detecting Internal Pressure

This pressure switch detects the pressure inside the electronics section. The pressure switch is
installed on the protection system and set to turn off at 392 ±20 Pa.
(1) Stop the operation (See “3.3.4 Stopping operation” for this procedure).
(2) Stop the supply of power.
(3) Confi rm that the LED (Green) of “POWER” is OFF.

CAUTION

The isothermal oven is extremely hot after turning off the power immediately. Keep the purging
air supplied for more than an hour after turning off the power. Keep hands away from the oven
components.

(4) Turn off the purging air.
(5) Loosen the locknuts on the cover for the protection system A of the right side and remove

the cover.