Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 87

Advertising
background image

<1. Overview>

1-60

IM 11B08A01-01E

Users at Level C or higher can cancel the calibration/validation command until the operation

mode changes to the stop mode.

Issue a command to start calibration (or validation) after confi rming that manual

replacement of the fl ow with the sample has been completed. When calibration (or
validation) is completed for all the specifi ed streams, the operation mode returns to the stop
mode in the stream specifi cation status.

Run

Stop

Stop

Run

W7

W7

W8

S8

W1

W1

S1

S7

S7

Time

Stream specification 1

Calibration 1

(Replaceable)

(Replaceable) Stream specification 1

Assinning calibration 1

Start of calibration (or validation)

End of calibration (or validation)

● In the case where a (manual) calibration/validation command is received during the stop

mode in the stream specifi cation status

The measurement status immediately changes to “Replaceable” in the stop mode in the

requested calibration (or validation) status.

Issue a command to start calibration (or validation) after confi rming that manual

replacement of the fl ow with the sample has been completed. When calibration (or
validation) is completed for all the specifi ed streams, the operation mode returns to the stop
mode in the stream specifi cation status.

Stop

Stop

Run

W7

W7

W8

S8

S7

S7

Time

Stream specification 1

Calibration 1

(Replaceable)

(Replaceable) Stream specification 1

Assinning calibration 1 Calibration (or validation) start command

Calibration (or validation) stop command

● In the case where a (manual) calibration/validation command is received during the pause

mode in the stream specifi cation status

The measurement status immediately changes to “Replaceable” in the stop mode in the

requested calibration (or validation) status.

Issue a command to start calibration (or validation) after confi rming that manual

replacement of the fl ow with the sample has been completed. When calibration (or
validation) is completed for all the specifi ed streams, the operation mode changes to the
stop mode in the stream specifi cation status.

Stop

Pause

Stop

Run

W7

W7

W8

S8

S7

S7

Time

Stream specification 1

Calibration 1

(Replaceable)

(Replaceable) Stream specification 1

Assinning calibration 1 Calibration (or validation) start command

Calibration (or validation) stop command

● In the case where a stop command is received during the run mode in the (manual)

calibration/validation status [4E2]

The operation mode is immediately prepared to be changed from the run mode to the stop

mode in the calibration (or validation) status.

The measurement status changes to “Replaceable” in the stop mode in the calibration (or

validation) status when the current measurement is completed.

The stop command can be canceled with a cancel command until the operation mode

changes to the stop mode.

2nd Edition : May 11, 2012-00