App.b-2 – Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 435

Advertising
background image

<Appendix B Terminology>

App.B-2

IM 11B08A01-01E

Term

Description

Icon/abbr.

SYS

SYS is the minimum analysis unit. Multiple SYSs can

be set in each GCM. SYSs in the same GCM are

synchronized for analysis, while those in different GCMs

are not synchronized.

Up to two SYSs can be set in an isothermal oven.

Up to two for one isothermal oven

Up to four for two isothermal ovens

Up to six for three isothermal ovens

Valves (Valve, RV, or LSV), detectors, and EPCs are set

for each SYS. Their respective operations are set in the

SYS method. Although atmospheric-pressure balancing

valves (AtmV) belong to GCM, their operations are also

set in the SYS method. The operations of timing signals

are set by using DO in this method as well.

(SYS analysis cycle) = (GCM analysis cycle) / n

(n:

1-8)

GCM Method

Any of methods No. 1 to No. 6 can be set for each GCM.

In the Manual status or Process–Stop status, the method

number will be blank.

Main Cycle

Time from start (0 second) to stop of analysis

Warming Up Time Time for displacement in sample streams
Stream

Valve

ON/OFF Time

ON/OFF time of sample stream valves

Pause Time

Time for measurement pause

SYS Method

Analysis Cycle

Time from start (0 second) to stop of analysis

Peak

Detection

Stop Time

Time to forcibly stop peak detection of chromatogram

Tracking

Specifi cation

Automatic

Renewal of

Tracking Factors

Valve

ON/OFF

Time

Valve

Atmospheric-

Pressure

Balancing Valve

ON/OFF Time

AtmV

DO

Operation

Setting

The operation setting of timing signals by using DO

EPC

Program

Setting

Purging

Displacement of the gas in the pressurized enclosures

with a protection gas (instrumentation air)

Custom Software

Capability

With the programming in YM-BASIC (Yokogawa’s original

programming language based on BASIC), calculation

formulas for analysis results and various measurement

statuses such as stream switching can be changed. This

function also allows special calculations using the data of

analog inputs from other analyzers and/or contact inputs.

Gate Tracking

Function

The automatic peak tracking function of GC1000 Mark

II was improved. The setting of tracking (correction)

conditions of the gate time setting to detect peaks has

become fl exible, from simultaneously setting all GCs to

setting each SYS or each detector.

This enables precise measurement even in

multicomponent analysis such as PIONA.

2nd Edition : May 11, 2012-00