<2. Installation, Piping, and Wiring>

2-17

IM 11B08A01-01E

CAUTION

• The unused electrical connection ports should be closed with an appropriate fl ameproof-

certifi ed plug.

• The blind plug shall not be used with an adapter.
• Analyzers have pressurized enclosures. The cable end should be sealed in order to

apply pressure to the pressurized enclosure. Otherwise, power does not supplied to the
electronics section.

M25x1.5

8 mm

32 mm

9/16 inch

A

3/4NPT

M25x1.5

G3/4

Connection port of the protection system

Connection port of
the electronics section *1

Electronic section
(Pressurized enclosure)

Protection system

(Flameproof enclosure)

Plug *2

Plate

Gasket

Nut

Connector-2

*1: Any of the six connection ports of the electronics

section can be equally used.

*2: A plug is attached to connector 2. Remove and keep it

for future use.

*3: Connector 1 is not attached at shipment.

One of the three types of connector 2 is attached at

shipment.

“N” is inscribed on the 3/4NPT connector and “G”

on the G3/4 connector.

(No inscription on the M25x1.5 connector)

Detail of A (Connector)

Connector-2

Connector-1 (accessory)

Figure 2.16

Wiring in ATEX or IECEx specifi cation

Connection port for the protection system

Remove the attached plug and perform wiring.
For the 3/4NPT or G3/4 connection port, use the connectors in the accessory kit.

Connection port for the electronics section

Six connection ports are provided in the electronics section. Use convenient ones.
Remove the attached plug of the connector and perform wiring.