2 procedure for replacing parts, Precautions for parts replacement – Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 356

Advertising
background image

<6. Maintenance>

6-3

IM 11B08A01-01E

Table 6.3

Recommended Parts List for Periodical Replacement

Item

Name

Part Number

Replacement

Interval

Note

Rotary Valve

Valve Seat

K9409ZA

100,000 cycles

CSW (Rulon) (2xK9034DN)

K9034FF

0.33 μL (Rulon)

K9409ZE

1 μL (Rulon) (2xK9034DS)

K9409ZD

2 μL (Rulon) (2xK9034DR)

K9409ZC

3 μL (Rulon) (2xK9034DQ)

K9409ZB

Over 10 μL (Rulon) (2xK9034DP)

K9034FR

CSW (Tefl on)

K9402XX

0.33 μL (Tefl on)

K9034FL

1 μL (Tefl on)

K9034FK

2 μL (Tefl on)

K9034FJ

3 μL (Tefl on)

K9034FH

Over 10 μL (Tefl on)

Valve Gland

K9034BA

5-years

CSW (SUS316)

K9034BP

0.33 to 3 μL (SUS316)

K9034BJ

10 μL (SUS316)

K9034BH

25 μL (SUS316)

K9034BG

50 μL (SUS316)

K9034BB

360 μL or more (SUS316)

K9034GX

CSW (Hastelloy C)

K9034GV

0.33 to 3 μL (Hastelloy C)

K9034GS

25 μL (Hastelloy C)

K9034GR

50 μL (Hastelloy C)

K9034GN

360 μL or more (Hastelloy C)

Liquid Sampling

Valve

Seal Element

K9402VG

50,000 cycles

Rulon

K9402VH

Tefl on

O-ring

K9402UN

2-years

Filter

Element

L9862AG

1-year

Gasket

L9862AC

1-year

L9862AD

Stream Switching

Valve

Diaphragm

K9192WL

3-years

Viton + Tefl on

K9192WV

Silicone + Tefl on

Oven

Heater

K9409GE

5-years

Isothermal Oven, 100 V, 110 °C or below

K9409GN

Isothermal Oven, 100 V, over 110 °C

K9409GF

Isothermal Oven, 110 V, 110 °C or below

K9409GP

Isothermal Oven, 110 V, over 110 °C

K9409GG

Isothermal Oven, 115 V, 110 °C or below

K9409GB

Isothermal Oven, 115 V, over 110 °C

K9409GH

Isothermal Oven, 120 V, 110 °C or below

K9409GC

Isothermal Oven, 120 V, over 110 °C

K9409GS

Isothermal Oven, 200 V

K9409GT

Isothermal Oven, 220 V

K9409GU

Isothermal Oven, 230 V

K9409GV

Isothermal Oven, 240 V

Turbine

K9400FA

5-years

Packing

K9803SX

5-years

For large isothermal oven

K9803SW

For standard isothermal oven

Vortex tube

K9803HE

5-years

For FPD

6.2

Procedure for Replacing Parts

Precautions for Parts Replacement

(1) Always observe the instructions in this manual when replacing parts.
(2) When disconnecting pipes, do not allow any oil or moisture to get inside the pipes.

2nd Edition : May 11, 2012-00

6

Advertising