2 alarm process, 3 alarm number – Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 415

Advertising
background image

<7. Troubleshooting>

7-2

IM 11B08A01-01E

Alarm Level
Set to level 2 or 3.
Alarm Message
The alarm message can be defi ned with up to 22 alphanumeric characters.

7.1.2 Alarm

Process

There are four alarm levels which are processed as follows.

Level 1

This alarm is activated when malfunctions in the system or hardware occur. Once activated, the
alarm status is maintained until released.
When a level 1 alarm is activated during Process status with Run mode, Run mode moves to
Stop at the end of the measurement being performed at the time.

Level 2

This alarm is activated when malfunctions in measurement conditions, etc. occur.
Once activated, the alarm status is maintained until released.

Component

This alarm is activated when measurement results for concentration, retention time, etc. fall out
of range.
Once activated, the alarm status is maintained until released.

Level 3

This alarm is activated for minor malfunctions or information.
The alarm status is not maintained.

7.1.3 Alarm

Number

The alarm number is assigned for each alarm level as follows.

Alarm 1 to 174: Level 1 Alarm

1 to 174: Fixed Alarm

Alarm 201 to 294: Level 2 Alarm

201 to 232: D/I Alarm (200 + contact number)
251 to 294: Fixed Alarm
291 to 294: Composition Alarm

Alarm 401 to 562: Level 3 Alarm

401 to 432: D/I Alarm (400 + contact number)
450 to 562: Fixed Alarm
471 to 474: Composition Alarm

2nd Edition : May 11, 2012-00