<Appendix C Standard Specifi cations>

App.C-3

IM 11B08A01-01E

Purity:

Measuring range from 0 to 50 ppm or more:

99.99% minimum (water: 10 ppm or

less, organic components: 5 ppm or

less)

Measuring range from 0 to less than 50 ppm:

99.999% minimum (water: 5 ppm or

less, organic components: 0.1 ppm

or less)

Pressure:

H

2

:

500 kPa (72.5 psi) (supplied with

extra-regulator for explosion-proof

certifi cation)

Other than H

2

:

400 to 700 kPa

Consumption:

60 to 300 mL/min per isothermal oven

Connection: 6 mm or 1/4″ tube

3.2.3

Combustion air for FID/FPD:

Purity:

Measuring range from 0 to 50 ppm or more:

water: 10 ppm or less,

organic components: 5 ppm or less

Measuring range from 0 to less than 50 ppm:

water: 5 ppm or less,

organic components: 0.1 ppm or less

Pressure: 400 to 700 kPa

Consumption:

Approximately 300 mL/min per

detector

Connection: 6 mm or 1/4″ tube

4. Input and Output Specifi cations

Wiring method:

FM:

Conduit wiring (3/4NPT(F))

ATEX, IECEx:

Cable packing (G3/4(F),

3/4NPT(F), M25x1.5(F))

TIIS:

Cable packing (G3/4(F), 3/4NPT(F)),

Sealing fi tting (only for Ethernet

cable)

Note: Cable packing for TIIS is provided by Yokogawa.

Other conduit wiring or cable packing should be
prepared by the user.

Wiring connection:

FM-X, ATEX, IECEx, TIIS:

Contact outputs for System Alarm 1,

Annunciator: Explosion

proof

enclosure

Other I/Os: Control unit

FM-Y:

All I/Os: Control unit

4.1 Communication
4.1.1

Connection to Analyzer network

Included as standard.

Communication standard:

Ethernet

Connection type:

IEEE802.3U

100Base-TX (RJ-45 shielded twisted

pair cable) or 100Base-FX (SC fi ber-

optics cable)

Channel: 1 or 2

Protocol: TCP/IP, FTP

Data to be transmitted:

Analysis results, calibration factors,

alarms, status, and chromatogram

Data to be received:

Operation requests (stream sequence

setting, stream setting, run, stop,

pause, and range change)

Device to be connected:

PCAS, ASET, ASGW, GC8000 (LCD),

and OPC through FCN/FCJ

Note: SHDSL interface (K9802PB) is provided as an

option (1 channel only).

External I/O Cutoff Output:

Number of outputs: 2

Function: Monitoring the purge air pressure

in the electronics section, applying

power (24 V DC) to the signal

interrupter when the state is normal.

Signal interrupter (Rack-mounted type:

K9806AA, Desktop type:K9806AB):

FM-X, ATEX, IECEx, TIIS:

Additionally required (only with

twisted pair cables). Signal is

interrupted by power supply OFF

signals from the external I/O cutoff

output.

FM-Y: Not

required

Note: For installation in hazardous area, an explosionproof

structure authenticated by a relevant certifi ed body
shall be prepared by customer.

4.1.2

Connection to DCS

Both Ethernet and Serial communication are

available.

(1) Ethernet communication

The same port as analyzer network is used.

Refer to 4.1.1 Connection to Analyzer network.

Communication standard:

Ethernet

Protocol: Modbus/TCP

Communication speed:

100Mbps

Number of DCS connections: Maximum of 4

(2) Serial communication (Option)

Channnel:

1 or 2

Communication standard: RS-422

Transmission: Full duplex (4-wire system)

Protocol: Modbus, Y-Protocol (GC1000/GC8,

GC6 and BTU for Japan)

Note: Concurrent usage of Y-Protocol is not available.

Start-stop (asynchronous) communication:

Start bit 1, Data bits 7 (ASCII)/Data

bits 8 (RTU), Parity bit 1, Stop bit 1

Parity check: Odd/even/none

Communication speed:

1200/2400/4800/9600/19200/38400

bps

Transmission mode:

ASCII or RTU

Note: Only the ASCII format is used in the Y-Protocol.

Communication control:

None/handshake

Data to be transmitted:

Analysis results, calibration factors,

and alarms

2nd Edition : May 11, 2012-00