<2. Installation, Piping, and Wiring>

2-14

IM 11B08A01-01E

2.3.2 Recommended

Cables

Table 2.4

Wiring

Connections

Cable

Inlet

Cable

O.D.

(for TIIS)

Wiring

Cable Condition

Terminal

Cable

Shield

Protection system

A (right): 2

B (left): 1

ø8.0 to

16.0 mm

Protection system A:

(A) Electric circuit and

heater power

Protection system B:

(B) Heater power

3.5 to 5.5 mm

2

max.

1.25 to 5.5 mm

2

max.

M4 screw

crimp-on

terminal

Not

required

(C) Contact output for

system alarm 1

(D) Contact output for

annunciator

0.75 to 1.5 mm

2

max.

Cable length 1 km max.

For

MKKDSN

(Note 2)

Required

Electronics section

6

(Junction

box as

needed)

ø9.0 to

16.0 mm

Without protection

system:

(A) Electric circuit and

heater power

(B) Heater power

3.5 to 5.5 mm

2

max.

1.25 to 5.5 mm

2

max.

M4 screw

crimp-on

terminal

Not

required

Without protection

system:

(C) Contact output for

system alarm 1

(D) Contact output for

annunciator

0.75 to 1.5 mm

2

max.

Cable length 1 km max.

For FKC

(Note 2)

Required

(E) Analog input

(16 points max.)

(F) Contact input

(32 points max.)

(G) Contact output

(Note 1)

(20 points max.)

(H) Serial

communication

(J) Analog output

(32 points max.)

0.5 to 1.5 mm

2

max.

Cable length 1 km max.

For FKC

(Note 2)

Required

(K) Grounding

5.5 mm

2

or more

Grounding resistance of

100 ohms max.

M4 screw

crimp-on

terminal

Not

required

(L) External I/O cutoff

output (Power

cutoff signal)

0.75 to 1.5 mm

2

max.

Cable length 1 km max.

Twisted-pair cable

For FKC

(Note 2)

Required

NA

(use

sealing

fi tting)

(M) Ethernet (shielded

twisted-pair cable)

CAT.5/CAT.5E

50 m or less

RJ45

Required

(N) Ethernet (fi ber-

optic cable)

For 1300 nm Outdoor

type

multi-mode of 50/125 μm

or 62.5/125 μm

SC

Note 1: Use double-isolation cables for the contact output line (AC).

Double-isolate either contact output line (AC) or (DC) if they are mixed.

Note 2: Use MKKDSN series terminals (manufactured by Phoenix Contact Ltd.) for the protection system, and

FKC series terminals (manufactured by Phoenix Contact Ltd.) for the electric circuit except for the power
or Ethernet line. For these wiring connections, use AI series crimp-on terminals manufactured by the
same company. Four types of crimp-on terminals are used according to the wire diameters (see Table
2.5).

Table 2.5

Crimp-on terminals

Terminal Series

Cable Core

Cable O.D.

Terminal Type Peel off length

MKKDSN

0.75 mm

2

Less than ø2.8 mm

AI 0.75-6GY

Approx. 6 mm

1 mm

2

Less than ø3.0 mm

AI 1-6RD

1.5 mm

2

Less than ø3.4 mm

AI 1.5-6BK

FKC

0.5 mm

2

Less than ø2.5 mm

AI 0.5-10WH

Approx. 10 mm

0.75 mm

2

Less than ø2.8 mm

AI 0.75-10GY

1 mm

2

Less than ø3.0 mm

AI 1-10RD

1.5 mm

2

Less than ø3.4 mm

AI 1.5-10BK

Contact Phoenix Contact Ltd. for details.

2nd Edition : May 11, 2012-00