5.4.12 Alarm

Setup

1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5

Alarm Setup

GCM # 1 Set # 1
>

- Stream number 2

Peak number 1
- Check item Conc
- Maximum 0.000
- Minimum 0.000

Menu

GCM#

Set#

F1

F2

F3

F4

F5

F6

Figure 5.86

Example of alarm setup screen

F1 (Menu):

Displays the Table Menu screen.

F5 (GCM #): Specifi es a GCM number.
F6 (Set #):

Sets an alarm number.

• When a stream and a peak number in database are both zero, they are assessed as no

settings and all data is displayed with * (asterisk).

Display Item

Name

Set Lower Limit Upper Limit

Unit

Remarks

GCM #
Set #

Relevant set number (1 to 32),

32 per GCM (Total 192)

Stream

number

(0), 1

31

99 can also be entered.

Peak number

(0), 1

999

When a stream number is 99,

all streams belonging to GCM is

applicable and a relative peak

number is assigned to the peak

number.

Assign 0 to the stream number

and the peak number to enable

setting invalid data.

Check items

Alarm check items, Density, R.

Time, Vari coeff, Tailing coeff

Upper Limit,

Lower Limit

Density: −999.999 to 9999.999

(Process assign is A/I signal),

0.000 to 9999.999 (Process

assign is other than A/I signal)

R. Time:0.000 to 21600.0 sec

Vari coeff: 0.000 to 1.999

Tailing coeff: 0.000 to100.000