<6. Maintenance>

6-31

IM 11B08A01-01E

(3) Turn off the supplies of the purging air, carrier and sample gas.
(4) Open the control unit door using a key included in the accessory kit.
(5) Remove the two screws to remove the TCD cover.
(6) Remove the insulator.
(7) Remove the four couplings connected to the TCD. When removing the couplings, label

them for ease of identifying the connections later.

(8) Remove the four wires connected to the TCD terminals.

IMPORTANT

The direction of the TCD depends on the type of carrier gas. See the inscription on the TCD.

Carrier gas

Inscription

H

2

, He

H N

N

2

, Ar

N H

(9) Remove the two screws which fi x the TCD.
(10) Install a new TCD.
(11) Reconnect the wires and pipes in the reverse order of steps (4) to (9). Confi rm wiring and

perform a leak test.

High-temp. type TCD

(1) Stop the operation. (See “3.3.4 Stopping operation” for this procedure).
(2) Stop the supply of power.

CAUTION

The isothermal oven is extremely hot after turning off the power immediately. Keep the purging
air supplied for more than an hour after turning off the power. Keep hands away from the oven
components.

(3) Turn off the supplies of the purging air, carrier and sample gas.
(4) Open the control unit door using a key included in the accessory kit.
(5) Remove the two screws to remove the TCD cover.
(6) Remove the insulator.
(7) Remove the wires connected to the TCD card from the connectors, and pull into the

isothermal oven through the bushing holes.

(8) Remove the six screws from the cover of the electronics section in the oven unit by using a

hex wrench (an accessory) and open the cover.

(9) Remove the four couplings connected to the TCD. When removing the couplings, label

them for ease of identifying the connections later.

(10) Remove the two screws which fi x the TCD.
(11) Install a new TCD.
(12) Reconnect the wires and pipes in the reverse order of steps (4) to (8). Confi rm wiring and

perform a leak test.

<Replacement Parts>

Low temp. type K9192QA

For normal usage (Aluminum)

K9192QB

Corrosion-resistant type (SUS316)

K9192QJ

High-sensitivity type (Aluminum)

2nd Edition : May 11, 2012-00

6