<Appendix E Computation Scheme>

App.E-8

IM 11B08A01-01E

• The analysis elapse time is greater than the gate-on time.
• The previous peak is ended.
• The detector signal exceeds the detection level.
• The gradient of the detector signal (change with time) exceeds the peak-on detection slope

value.

When all the following conditions are satisfi ed, it is assumed that the peak is ended (off):

• The analysis elapse time passes the gate-off time.
• The detector signal passes the peak point.
• The detector signal is lower than the detection level.
• The gradient of the detector signal (change with time) is smaller than the peak-off detection

slope value.

If the next gate is started before the detector signal passes the peak point, it is also assumed that

the peak is ended (off).
The detection level can be set for each peak based on the signal level upon the start of analysis

(or the base level when the base level setting is specifi ed).
The peak-on detection slope value is equal to the detection slope value (set for each peak).
The peak-off detection slope value is the detection slope value multiplied by the detection slope

on/off ratio (that can be group-set).

F0205020104.ai

Peak-on detection
slope value

Peak-on detection slope value

Base level

Peak

gate

(e) Zone Gating

Zone gating method detects peaks under the upper limitation of specifi ed number of peaks in the

range (zone) between the gate-on time and the gate-off time.

(i) Peak detection action in zone gating

The peak detection action in the zone gating acquires the number of peaks assigned to the zone

using the gate on as a trigger and like the regular slope detection, starts the detection of peak on.

As soon as the peak on is detected, the detection of peak off is started. At the point the peak off is

detected, the fi rst peak is determined.
The following peak detection is started and then the second and third peaks are determined. This

peak detection action is continued until the peaks of assigned number are determined. After the

detection of the fi nal peak in the zone has been completed, the peak detection is not performed

until the gate-off time.

2nd Edition : May 11, 2012-00