<6. Maintenance>

6-22

IM 11B08A01-01E

(5) Remove the couplings connected to the inlet (IN) and outlet (OUT) of the regulator. See

Figure 6.19. Label the couplings with “IN” and “OUT” for ease of identifi cation later.

Note

Remove any metal fl akes off the couplings.

(6) Remove the stem lock.
(7) Remove the mounting locknut and the regulator.
(8) Install a new regulator in the reverse order of steps (5) through (7).
(9) Start supplying the carrier gas and other gases and perform a leak test around the

couplings.

(10) Set the carrier gas pressure to the value indicated in the operation data. Adjust and set the

pressure by loosening the stem lock nut and rotating the stem. After completing the setting,
fi x the stem with the locknut.

(11) Close the isothermal oven door and the cover of the pressure/fl ow control section.

IMPORTANT

After the temperature of the isothermal oven in the oven unit becomes substantially stable, set
the pressure of the carrier gas again according to the Operation data.

<Replacement Parts>

Applicable Temperature Range O-ring Material

K9409AW

180 °C or less

Viton

K9409AV

225 °C or less

Kalrez

IN

OUT

Isothermal Oven Door

Regulator for
Carrier Gas

Regulator for Carrier Gas

Outside Wall of

the Pressure Control Chamber

Inside Wall of

the Isothermal Oven

Nut

Stem Lock

Couplings

Figure 6.19

Regulator for carrier gas

2nd Edition : May 11, 2012-00