Setup Menu

Model 6200T Total Sulfur Analyzer

Teledyne Analytical Instruments

100

5.4.3. Reporting Range Modes: Single, Dual, Auto Ranges

The 6200T provides three analog output range modes to choose from:

 Single range (SNGL) mode sets a single maximum range for the analog

output. If single range is selected (refer to Section 5.4.3.1) both outputs are
slaved together and will represent the same measurement span (e.g. 0-50
ppm), however their electronic signal levels may be configured for different
ranges (e.g. 0-10 VDC vs. 0-.1 VDC – Refer to Section 5.9.3.1).

 Dual range (DUAL) allows the A1 and A2 outputs to be configured with

different measurement spans (refer to Section 5.4.3.2).

 Auto range (AUTO) mode gives the analyzer to ability to output data via a

low range and high range. When this mode is selected (refer to Section
5.4.3.3) the 6200T will automatically switch between the two ranges
dynamically as the concentration value fluctuates.

Also, in this mode the RANGE Test function displayed on the front panel during
SAMPLE mode will be replaced by two separate functions, Range1 and Range2.

Range status is also output via the External Digital I/O Status Bits (refer to Section
8.1.1).
To select the Analog Output Range Type press:

SAMPLE RANGE = 500.000 PPB SO2 =XXX.X

< TST TST > CAL

SETUP

SAMPLE

ENTER SETUP PASS : 818

8 1 8

ENTR

EXIT

SETUP X.X

CFG DAS

RNGE

PASS CLK MORE

EXIT

SETUP X.X

RANGE CONTROL MENU

MODE

SET UNIT

EXIT

SETUP X.X

RANGE MODE:

SNGL

SNGL

DUAL AUTO

ENTR EXIT

EXIT

Returns

to the Main

SAMPLE

Display

Go To

Section

6.7.4

Go To

Section

6.7.5

Go To

Section

6.7.6

Only one of the

range modes may

be active at any

time.

Figure 5-3: SETUP RNGE – Reporting Range Mode