Manual calibration with izs option – Teledyne 6200T - Sulfides Analyzer User Manual

Page 197

Advertising
background image

6200T Total Sulfur Analyzer

Calibration Procedures

Teledyne Analytical Instruments

197

9.5. Manual Calibration with IZS Option

Under the best conditions, the accuracy off the SO

2

effusion rate of the IZS option’s

permeation tube is about ±5%. This can be subject to significant amounts of drift as the
tube ages and the amount of SO

2

contained in the tube is depleted. Whereas this may be

sufficient for informal calibration checks, it is not adequate for formal calibrations and is
not approved for use by the US EPA as a calibration source.
Therefore, for formal calibrations of an instrument with an IZS option installed the
following provisions must be followed.

 Zero air and span gas must be supplied to the analyzer through the sample

gas inlet as depicted in

 Figure 9-1 of Section 9.2.
 The calibration procedure must be initiated using the CAL button, not the

CALZ or CALS buttons, using the procedure defined in Section 9.2.

 Using the CAL button does not activate the zero/span or sample/cal valves

of the IZS option, thus allowing the introduction of zero air and sample gas
through the sample port from more accurate, external sources such as a
calibrated bottle of SO

2

or a Model T700 Dilution Calibrator.

SAMPLE RANGE = 500.000 PPB SO2 =XXX.X


< TST TST >

CAL CALZ CALS

SETUP

Use for formal

calibration

operations.

Use only for

informal calibration

checks.

Figure 9-8: Manual Calibration with IZS Option

9.6. Manual Calibration Checks with IZS or Zero/Span Valves

Zero and span checks using the zero/span valve or IZS option are similar to that
described in Section 9.3, with the following exceptions:

 On units with an IZS option installed, zero air and span gas are supplied to

the analyzer through the zero gas inlet and from ambient air.

 On units with a zero/span valve option installed, zero air and span gas are

supplied to the analyzer through the zero gas and span gas inlets from two
different sources.

 The zero and calibration operations are initiated directly and independently

with dedicated buttons CALZ and CALS.

Advertising