Technical assistance – Teledyne 6200T - Sulfides Analyzer User Manual
Page 284
Troubleshooting
Model 6200T Total Sulfur Analyzer
284
Teledyne Analytical Instruments
QUESTION ANSWER
How often do I need to change the
particulate filter?
Once per week. Table 11-1 contains a maintenance schedule listing the most
important, regular maintenance tasks.
What is the averaging time for an
6200T?
The default averaging time, optimized for ambient pollution monitoring, is 240
seconds for stable concentrations and 20 seconds for rapidly changing
concentrations; Refer to 13.7.1 for more information.
My analyzer has the optional, user -
configurable analog output channels.
How do I program and use them?
Instructions for this can be found in the Manual Addendum for Configurable Analog
Output, PN 06270.
How long does the sample pump last? The sample pump should last about one year and the pump diaphragms should to
be replaced annually or when necessary. Use the PRES test function displayed via
the front panel to see if the diaphragm needs replacement (refer to Section 12.1.2).
Do I need a strip chart recorder or
external data logger?
No, the 6200T is equipped with a very powerful internal data acquisition system.
Section 6.6 describes the setup and operation in detail.
12.10. Technical Assistance
If this manual and its troubleshooting / repair sections do not solve your problems,
technical assistance may be obtained from:
Teledyne Analytical Instruments, Customer Service,
16830 Chestnut Street
City of Industry, California 91748, USA
TEL: 626-934-1500 or 888-789-8168
FAX: 626-934-1651
EMAIL:
TETCI_
customerservice @teledyne.com
www.teledyne-ai.com
Before you contact Teledyne Analytical Instruments’ Customer service, fill out the
problem report form in Appendix C, which is also available online for electronic
submission at
http://www.teledyne-ai.com