5.9.2. Analog Output Step Test

Analog Output is used as a step test to check the accuracy and proper operation of the
analog outputs. The test forces all four analog output channels to produce signals
ranging from 0% to 100% of the full scale range in 20% increments. This test is useful
to verify the operation of the data logging/recording devices attached to the analyzer.
Access the Analog Output Step Test from the DIAG Menu (refer to Figure 5-16), then
press:



SAMPLE RANGE = 500.000 PPB SO2 =XXX.X

< TST TST > CAL

SETUP

SAMPLE

ENTER SETUP PASS : 818

8 1 8

ENTR

EXIT

SETUP X.X

PRIMARY SETUP MENU

CFG DAS

RNGE

PASS CLK

MORE

EXIT

DIAG

SIGNAL I / O

NEXT

ENTR

EXIT

Exit-Exit

returns to the

DIAG

menu

SETUP X.X

SECONDARY SETUP MENU

COMM VARS

DIAG

EXIT

DIAG

ANALOG OUTPUT

PREV

NEXT

ENTR

EXIT

DIAG AOUT

ANALOG

OUTPUT

0%

EXIT

Performs

analog output

step test.

0% - 100%

DIAG AOUT

ANALOG OUTPUT

[0%]

EXIT

Pressing the “0%” button while performing the test will

pause the test at that level. Brackets will appear around
the value: example: [20%] Pressing the same key again

will resume the test.