6200T Total Sulfur Analyzer

Setup Menu

Teledyne Analytical Instruments

119

5.9.3. Analog I/O Configuration

Table 6-8 lists the analog I/O functions that are available in the 6200T.

Table 5-4: DIAG - Analog I/O Functions

SUB MENU

FUNCTION

AOUTS CALIBRATED:

Shows the status of the analog output calibration (YES/NO) and initiates a calibration of all
analog output channels.

CONC_OUT_1

Sets the basic electronic configuration of the A1 analog output (SO

2

). There are three options:

RANGE:

Selects the signal type (voltage or current loop) and full scale level of the

output.

REC_OFS:

Allows setting a voltage offset (not available when RANGE is set to Current

Loop (CURR).

AUTO_CAL:

Performs the same calibration as AOUT CALIBRATED, but on this one

channel only.

NOTE: Any change to RANGE or REC_OFS requires recalibration of this output.

CONC_OUT_2

Same as for CONC_OUT_1 but for analog channel 2 (SO

2

)

TEST OUTPUT

Same as for CONC_OUT_1 but for analog channel 3 (TEST)

CONC_OUT_3

(Not available in the analyzer’s standard configuration; applies when optional sensor installed).

AIN CALIBRATED

Shows the calibration status (YES/NO) and initiates a calibration of the analog input channels.

XIN1

.
.
.

XIN8

For each of 8 external analog inputs channels, shows the gain, offset, engineering units, and
whether the channel is to show up as a Test function.

Table 5-5: Analog Output Voltage Ranges

RANGE

MINIMUM OUTPUT

MAXIMUM OUTPUT

0-0.1 V

-5 mV

+105 mV

0-1 V

-0.05 V

+1.05 V

0-5 V

-0.25 V

+5.25 V

0-10 V

-0.5 V

+10.5 V

The default offset for all ranges is 0 VDC.