5.4.5. Dilution Ratio (Option)

The dilution ratio is a software option that allows the user to compensate for any dilution
of the sample gas before it enters the sample inlet. Once the degree of dilution is known,
add an appropriate scaling factor to the analyzer’s SO

2

concentration calculation so that

the measurement range and concentration values reflect the undiluted values when
shown on the instrument’s front panel display screen and reported via the analog and
serial outputs.
Using the dilution ratio option is a 4-step process:

1. Select reporting range units: Follow the procedure in Section 5.4.4

2. Select the range: Use the procedures in Section 5.4. Ensure that the SPAN value

entered is the maximum expected concentration of the undiluted calibration gas
and that the span gas is either supplied through the same dilution inlet system as
the sample gas or has an appropriately lower actual concentration. For example,
with a dilution set to 100, a 1 ppm gas can be used to calibrate a 100 ppm
sample gas if the span gas is not routed through the dilution system. On the
other hand, if a 100 ppm span gas is used, it needs to pass through the same
dilution steps as the sample gas.

3. Set the dilution factor as a gain (e.g., a value of 20 means 20 parts diluent and 1

part of sample gas):

SAMPLE RANGE = 500.000 PPB SO2 =XXX.X

< TST TST > CAL

SETUP

SETUP

PRIMARY SETUP MENU

CFG DAS

RNGE

PASS CLK MORE

EXIT

SETUP

RANGE CONTROL MENU

MODE

SET UNIT

DIL

EXIT

DIL

only appears

if the dilution ratio

option has been

installed

SETUP

DIL FACTOR: 1.0 GAIN

0 0 0 1 .0

ENTR

EXIT

Toggle to set the dilution factor.

This is the number by which the

analyzer will multiply the SO

2

concentrations of the gas passing

through the reaction cell.

SETUP

DIL FACTOR: 20.0 GAIN

0 0 2 0 .0

ENTR

EXIT

EXIT

ignores the

new setting.

ENTR

accepts the

new setting.

SAMPLE

ENTER SETUP PASS : 818

8 1 8

ENTR

EXIT

Figure 5-8: SETUP RNGE – Dilution Ratio

The analyzer multiplies the measured gas concentrations with this dilution factor and
displays the result.