Communications Setup

Teledyne Analytical Instruments

145

Table 6-3: LAN/Internet Default Configuration Properties

PROPERTY DEFAULT

STATE

DESCRIPTION

DHCP

ON

This displays whether the DHCP is turned ON or OFF. Press
EDIT and toggle ON for automatic configuration after first
consulting network administrator. (

INSTRUMENT IP

ADDRESS

This string of four packets of 1 to 3 numbers each (e.g.
192.168.76.55.) is the address of the analyzer itself.

GATEWAY IP

ADDRESS

0.0.0.0

Can only be edited when DHCP is set to OFF.
A string of numbers very similar to the Instrument IP address
(e.g. 192.168.76.1.) that is the address of the computer used
by your LAN to access the Internet.

SUBNET MASK

0.0.0.0

Can only be edited when DHCP is set to OFF.
Also a string of four packets of 1 to 3 numbers each (e.g.
255.255.252.0) that identifies the LAN to which the device is
connected.
All addressable devices and computers on a LAN must have
the same subnet mask. Any transmissions sent to devices
with different subnets are assumed to be outside of the LAN
and are routed through the gateway computer onto the
Internet.

TCP PORT

1

3000

This number defines the terminal control port by which the
instrument is addressed by terminal emulation software, such
as Internet or Teledyne Analytical Instruments’s AICOM.

HOST NAME

6200T

The name by which your analyzer will appear when
addressed from other computers on the LAN or via the
Internet. To change, see Section 6.5.2.1.

1

Do not change the setting for this property unless instructed to by Teledyne Analytical Instruments’s

Customer Service personnel.