Figure 5-20: diag – analog output calibration mode – Teledyne 6200T - Sulfides Analyzer User Manual

Page 122

Advertising
background image

Setup Menu

Model 6200T Total Sulfur Analyzer

Teledyne Analytical Instruments

122

5.9.3.2. Analog Output Calibration Mode

Analog output calibration should to be carried out on first startup of the analyzer
(performed in the factory as part of the configuration process) or whenever recalibration
is required. The analog outputs can be calibrated automatically, either as a group or
individually, or adjusted manually.
In its default mode, the instrument is configured for automatic calibration of all
channels, which is useful for clearing any analog calibration warnings associated with
channels that will not be used or connected to any input or recording device, e.g.,
datalogger.
Manual calibration should be used for the 0.1V range or in cases where the outputs must
be closely matched to the characteristics of the recording device. Manual calibration
requires the AUTOCAL feature to be disabled.

To calibrate the outputs as a group, activate the ANALOG I/O CONFIGURATION
MENU from the DIAG Menu (refer to Figure 5-16), then press:

DIAG AIO

AUTO

CALIBRATING CONC_OUT_1

(continues for any active channels)

AUTO

CALIBRATING

TEST_OUTPUT

Exit at any time

to return to the

main DIAG

menu

DIAG

ANALOG I / O CONFIGURATION

PREV

NEXT

ENTR

EXIT

DIAG AIO

AOUTS CALIBRATED:

NO

< SET

SET>

CAL

EXIT

If AutoCal has been

turned off for any

channel, the message

for that channel will be

similar to:

NOT AUTO CAL

CONC_OUT_1

Exit to return to

the I/O

configuration

menu

If any of the channels have

not been calibrated this

message will read NO.

DIAG AIO AOUTS CALIBRATED:

YES

< SET

SET> CAL

EXIT

Figure 5-20:

DIAG – Analog Output Calibration Mode