6200T Total Sulfur Analyzer

Setup Menu

Teledyne Analytical Instruments

99

concentration of SO

2

is typically less than 500 ppb, the full scale of expected values is

only 0.25% of the instrument’s full 20,000 ppb measurement range. Unmodified, the
corresponding output signal would also be recorded across only 0.25% of the range of
the recording device.
The 6200T solves this problem by allowing the user to select a scaled reporting range
for the analog outputs that only includes that portion of the physical range relevant to the
specific application. Only the reporting range of the analog outputs is scaled, the
physical range of the analyzer and the readings displayed on the front panel remain
unaltered.
Applications where low concentrations of SO

2

are measured require greater sensitivity

and resolution than typically necessary for measurements of higher concentrations.
The 6200T solves this issue by using two hardware physical ranges that cover the
instrument’s entire 0 and 20,000 ppb measurement range: a 0 to 2,000 ppb physical
range for increased sensitivity and resolution when measuring very low SO

2

concentrations, and a 0 to 20,000 ppb physical range for measuring higher SO

2

concentrations. The analyzer’s software automatically selects which physical range is in
effect based on the analog output reporting range selected by the user.

 If the high end of the selected reporting range is  2,000 ppb. The low

physical range is selected.

 If the high end of the selected reporting range is  2,001 ppb. The high

physical range is selected.

Once properly calibrated, the analyzer’s front panel display will accurately report
concentrations along the entire span of its 0 and 20,000 ppb physical range regardless of
which reporting range has been selected for the analog outputs and which physical range
is being used by the instrument’s software.