Teledyne 6200T - Sulfides Analyzer User Manual
Page 42
Getting Started
Model 6200T Total Sulfur Analyzer
Teledyne Analytical Instruments
42
The analyzer’s front panel liquid crystal display screen includes touch control. Upon
analyzer start-up, the screen shows a splash screen and other initialization indicators
before the main display appears, similar to
Figure 3-2 above (may or may not display a Fault alarm). The LEDs on the display
screen indicate the Sample, Calibration and Fault states; also on the screen is the gas
concentration field (Conc), which displays real-time readouts for the primary gas and
for the secondary gas if installed. The display screen also shows what mode the analyzer
is currently in, as well as messages and data (Param). Along the bottom of the screen is
a row of touch control buttons; only those that are currently applicable will have a label.
Table 3-2 provides detailed information for each component of the screen.
ATTENTION
COULD DAMAGE INSTRUMENT AND VOID WARRANTY
Do not use hard-surfaced instruments such as pens to touch the control
buttons.
Table 3-2: Display Screen and Touch Control Description
Field
Description/Function
Status
LEDs indicating the states of Sample, Calibration and Fault, as follows:
Name Color
State Definition
SAMPLE Green
Off
On
Blinking
Unit is not operating in sample mode, DAS is disabled.
Sample Mode active; Front Panel Display being updated; DAS data
being stored.
Unit is operating in sample mode, front panel display being updated,
DAS hold-off mode is ON, DAS disabled
CAL Yellow
Off
On
Blinking
Auto Cal disabled
Auto Cal enabled
Unit is in calibration mode
FAULT Red
Off
Blinking
No warnings exist
Warnings exist
Conc
Displays the actual concentration of the sample gas currently being measured by the analyzer in the
currently selected units of measure
Mode
Displays the name of the analyzer’s current operating mode
Param
Displays a variety of informational messages such as warning messages, operational data, test function
values and response messages during interactive tasks.
Control Buttons Displays dynamic, context sensitive labels on each button, which is blank when inactive until applicable.
Figure 3-3 shows how the front panel display is mapped to the menu charts illustrated in
this manual. The Mode, Param (parameters), and Conc (gas concentration) fields in the
display screen are represented across the top row of each menu chart. The eight touch
control buttons along the bottom of the display screen are represented in the bottom
row of each menu chart.