Teledyne 6200T - Sulfides Analyzer User Manual

Page 274

Advertising
background image

Troubleshooting

Model 6200T Total Sulfur Analyzer

274

Teledyne Analytical Instruments

11. If the cold block or TEC is to be changed disconnect the TEC driver board from the

preamplifier board.

 Remove the cooler fan duct (4 screws on its side) including the driver

board.

 Disconnect the driver board from the TEC and set the sub-assembly aside.
 Remove the end plate with the cooling fins (4 screws) and slide out the

PMT cold block assembly, which contains the TEC.

 Unscrew the TEC from the cooling fins and the cold block and replace it

with a new unit.

12. Re-assemble the TEC subassembly in reverse order.

ATTENTION

COULD DAMAGE INSTRUMENT AND VOID WARRANTY

The thermo-electric cooler needs to be mounted flat to the heat sink. If
there is any significant gap, the TEC might burn out. Ensure to apply heat
sink paste before mounting it and tighten the screws evenly and cross-
wise.

 Ensure to use thermal grease between TEC and cooling fins as well as

between TEC and cold block.

 Align the side opening in the cold block with the hole in the PMT housing

where the sample Chamber attaches.

 Evenly tighten the long mounting screws for good thermal conductivity.

13. Re-insert the TEC subassembly. Ensure that the O-ring is placed properly and the

assembly is tightened evenly.

14. Re-insert the PMT/HVPS subassembly.

 Don’t forget the gasket between HVPS and PMT.
 Use new plastic screws to mount the PMT assembly on the PMT cold

block.

15. Insert the LED and thermistor into the cold block.

16. Replace the desiccant bags with five new desiccant bags.

17. Carefully replace the end plate.

 Ensure that the O-ring is properly in place. Improperly placed O-rings will

cause leaks, which – in turn – cause moisture to condense on the inside of
the cooler causing the HVPS to short out.

18. Reconnect the cables and the reaction cell

 Be sure to tighten these screws evenly.

19. Replace the sensor assembly into the chassis and fasten with four screws and

washers.

20. Perform a leak check the system.

21. Power up the analyzer and verify the basic operation of the analyzer using the

ETEST and OTEST features (refer to Section 6.9.5 and 6.9.6) or by measuring
calibrated zero and span gases.

Advertising