Calibration Procedures

Model 6200T Total Sulfur Analyzer

Teledyne Analytical Instruments

198

To perform a manual calibration check of an analyzer with a zero/span valve or IZS
Option installed, use the following method:

STEP ONE: Connect the sources of Zero Air and Span Gas as shown below.

SAMPLE

ZERO AIR

EXHAUST

Source of

SAMPLE Gas

VENT if input is pressurized

Internal Zero/Span Option (IZS) – Option 51A

Ambient

Air

External Zero

Air Scrubber

Filter

MODEL 701

Zero Air

Generator

Source of

SAMPLE Gas

VENT if input is pressurized

MODEL 700

Gas Dilution Calibrator

(with O

3

generator option)

VENT

SAMPLE

ZERO AIR

EXHAUST

SPAN 1

Calibrated

SO

2

gas

(At high

concentration)

VENT

Needle valve

to control flow

Chassis

Chassis

Figure 9-9: Setup for Manual Calibration Check with Z/S Valve or IZS Option (Step 1)