Model 6200T Total Sulfur Analyzer

Teledyne Analytical Instruments

60

DB9 connectors on the analyzer’s rear panel to connect the chain of
instruments, COM2 port is no longer available for separate RS-232 or
RS-485 operation.

ATTENTION

COULD DAMAGE INSTRUMENT AND VOID WARRANTY

Printed Circuit Assemblies (PCAs) are sensitive to electro-static
discharges too small to be felt by the human nervous system. Failure to
use ESD protection when working with electronic assemblies will void
the instrument warranty. Refer to Section 13 for more information on
preventing ESD damage.


To install shunt in the last analyzer:

1. With NO power to the instrument, remove its top cover and lay the rear panel

open for access to the multidrop PCA, which is seated on the CPU.

2. On the multidrop PCA’s JP2 connector, use the shunt provided to jumper Pins

21

 22 as indicated in (Figure 3-14).

Figure 3-14:

JP2 Pins 21-22 on RS-232-Multidrop PCA

Note: If you are adding an instrument to the end of a previously configured chain,
remove the shunt between Pins 21

 22 of JP2 on the Multidrop/LVDS PCA in the

instrument that was previously the last instrument in the chain.